The PrevEx Model SNR705 is an industrial flammability analyzer that measures flammable vapor concentrations from 0 to 100% LFL using a heated flame cell. Its fully heated pneumatics system operates up to 270°C (518°F) to prevent condensation, with a response time under 1 second. It mounts directly to process ductwork and uses compressed air for sample aspiration without pumps. Power is 120 VAC (230 VAC optional), with four redundant Form A safety relays and two Form A SPST relays for Service Needed and Air/Fuel cutoff. It provides two 4-20 mA analog outputs and an RS-485 Modbus port for communication. The flame cell is explosion proof Class I, Division 1; the enclosure is TYPE 12/13. It is FM approved standard, with FMc optional.
The Model SNR705 is an industrial strength assembly consisting of a heated flame cell and an integrated controller that continuously measures total flammable vapor concentrations from 0 to 100% of the Lower Flammable Limit (LFL) range. Optional modifications allow the SNR705 to sample low oxygen and inert atmospheres.
| Control Instruments Corp. | |
|---|---|
| Product Category | Gas Instruments |
| Product Number | PrevEx Model SNR705 |
| Product Name | Vapor Flammability Analyzer |
| Measurement Type | %LEL or %LFL |
| Application | Ambient Air Monitoring; Process Gas |
| Gas Types | Combustible; Hydrocarbons; (0-100% LFL); H2; (0-100% LFL); Methane (CH4) / Natural Gas; (0-100% LFL) |
| Electrical Outputs | Current; Switch or Relay |
| Display | Digital |