Description:
The CM1.00 Certification Master is used to check and certify the accuracy of the TruPosition Probe. If you have a calibration system and want to maintain files on your TruPosition Probes, then the CM1.00 Certification Master is the best and quickest way to check calibration.
CM1.00
The TruPosition Probe must:
be set to zero using the Zero Master.
be inserted into the Certification Master.
be rotated in the Master until the “A” axis reads “0.00.” the “B” axis should read “.50 +/-.01mm
be rotated in the Master until the “B” axis reads “0.00.” the “A” axis should read “.50 +/-.01mm