The F14844 is a triode-type MCP-based ion detector designed for miniature mass spectrometers operating at pressures up to 1 Pa. It features a 14.5 mm effective area, two MCPs with a 12 µm channel diameter and 8-degree bias angle, and a gain of at least 1 × 10⁶ at 1 Pa. The detector operates at 0 to 50 °C, with a maximum MCP IN-MCP OUT voltage of 2.0 kV and MCP OUT-ANODE voltage of 0.1 kV. Its electrical capacitance is 5.0 pF between MCP IN LEAD and ANODE LEAD, and 4.0 pF between MCP OUT LEAD and ANODE LEAD. The package is compact and demountable, with dimensions of 32 mm by 25.5 mm and a height of 5.41 mm. The dark count is no more than 3 s⁻¹·cm⁻², and the resistance ranges from 300 to 600 MΩ.
Next-generation ion detector ideal for miniature portable mass spectrometers operating at high pressure (1 Pa).
The F14844 is an MCP assembly with a unique triode structure*1 designed to vary the potential gradient and to suppress ion feedback*2 noise. This makes the F14844 a useful tool for portable desktop mass spectrometers in which it is difficult to install a large vacuum pumping system. The F14844 is a breakthrough ion detector that allows downsizing and cost reduction of the equipment.
*1: Structure consisting of three electrodes which are the MCP-out, mesh anode, and dynode.
*2: Ion feedback is the cause of unstable operation that occurs under low vacuum conditions
| Hamamatsu Photonics Europe | |
|---|---|
| Product Category | Microchannel Plates |
| Product Number | F14844 |
| Product Name | Circular MCP: Low vacuum operation, compact demountable package |