The 8585A1 is a shielded, coaxial open circuit termination for calibrating vector network analyzers, providing a nominal 180° phase offset over a limited frequency range. It features a BNC 75Ω connector and is characterized by a phase accuracy of ±1.0° from DC to 2.0 GHz, ±2.0° from 2.0 to 3.0 GHz, and ±6.0° from 3.0 to 12.0 GHz, with a minimum reflection coefficient of 0.98. The design uses a captivated dielectric rod to eliminate dependence on test port connector tolerances, ensuring high repeatability and consistent performance.
Shielded, coaxial opens are used in the calibration of vector network analyzers to provide a nominal 180° phase offset from a compatible reference short over a wide range of frequencies. At these frequencies, open circuit terminations are inherently imperfect. Shielding the open essentially eliminates radiation loss, but creates a residual frequency-sensitive capacitance. An accurate knowledge of the open’s effective capacitance is essential to an accurate calibration of the analyzer. Maury opens are characterized for effective capacitance versus frequency by means of a fourth order polynomial curve fit, and the nominal capacitance coefficients are provided with each unit.
| Maury Microwave | |
|---|---|
| Product Category | Calibration Instruments |
| Product Number | 8585A1 |
| Product Name | VNA Calibration Standard |
| Calibration Signal | Signal / Waveform Generator Output |